Replace the Multi Purpose Valve

Tools required
p/n
5023-0240
Parts required
p/n
0100-1259
01080-83202
5067-4237
Preparations
Remove all capillary connections from the Multi Purpose Valve.
1
Remove the clamp with the inline filter.
Agilent 1290 Infinity II Flexible Pump User Manual
Description
Hex driver, ¼", slitted
Description
Blank nut (plastic)
Blank nut (stainless steel)
Multi Purpose Valve Head
2
Unscrew the black union nut and remove the head of the
purge valve by pulling it to the front.
Maintenance

Replace the Multi Purpose Valve

7
133