Agilent Technologies Infinity II 1290 User Manual page 101

Flexible pump
Hide thumbs Also See for Infinity II 1290:
Table of Contents

Advertisement

4
Disconnect the MCGV cable (1), unscrew the two screws
(2, 3) and remove the valve.
6
Tighten the two screws (1, 2) and connect the cable to its
connector (3).
Next Steps:
8
Install the inlet weaver, see
9
Reconnect solvent tubes for channels A-D from the MCGV to the degasser outlets.
Agilent 1290 Infinity II Flexible Pump User Manual
"Replace the Inlet Weaver"
Replace the Multi-Channel Gradient Valve (MCGV)
5
Place the new MCGV into position.
N O T E
Make sure that channel A of the MCGV is put at the
bottom-right position.
7
Install the MCGV cover.
on page 90.
7
Maintenance
101