7
Maintenance
Pump Head Procedures
7
Close the tool by closing the 3 screws at the connection
ring.
Heat exchanger capillary
Tool handle
126
C A U T I O N
Damage to the pump head
Using a wrong torque will damage the pump head.
For handling the torque wrench, setting and applying
the right torque, consult the manual of your torque
wrench.
8
Using a torque key, which is included to the 1290 Infinity
Service Kit p/n 5067-4699, set 5 Nm and fix the central
alignment screw.
Central
alignment screw
5 Nm
Agilent 1290 Infinity II Flexible Pump User Manual