LAN Configuration
Automatic configuration with Bootp
5 Make these entries for the instrument:
6 Click OK.
7 Leave Edit BootP Addresses by pressing Close.
8 Exit BootP Settings by pressing OK.
9 After each modification of the BootP settings (i.e. EditBootPSettings) a
10 Power cycle the Instrument.
11 Use the PING utility to verify connectivity by opening a command
208
• MAC address
• Host name, Enter a Hostname of your choice.
The Host Name must begin with "alpha" characters (i.e. LC1260)
• IP address
• Comment (optional)
• Subnet mask
• Gateway address (optional)
The configuration information entered is saved in the Tab File.
stop or start of the BootP service is required for the BootP service to
accept changes. See
or
"Restarting the Agilent BootP Service"
OR
If you changed the IP address, power cycle the instrument for the
changes to take effect.
window and typing:
Ping 192.168.254.11 for example.
The Tab File is located at
C:\Documents and Settings\All Users\Application Data\Agilent\BootP\TabFile
"Stopping the Agilent BootP Service"
Agilent 1290 Infinity II Flexible Pump User Manual
on page 209
on page 210.