3
Pull the Inlet Weaver out of the Inlet Valve.
5
Fix the fitting of the new inlet weaver to the MCGV.
Agilent 1290 Infinity II Flexible Pump User Manual
4
Insert the new inlet weaver to the inlet valve. Fix the
weaver with the plastic screw.
6
Reconnect tubings between MCGV and degasser.
Maintenance
Replace the Inlet Weaver
7
91