Target pressure not reached for quaternary pump degasser

Error ID: 29221
The target pressure of the quaternary pump degasser has not been
reached within the expected time.
• Parameter: Pressure in mbar
Probable cause
1
2

Solvent counter exceeded limit

Error ID: 29146
The counter for the solvent volume has exceeded the limit, which has been
set in the user interface.
Probable cause
1
2
Agilent 1290 Infinity II Flexible Pump User Manual
Condensation in degasser chamber due to
temperature fluctuation.
Degasser is defect.
No solvent present.
Inappropriate setting.
Error Information
Pump Error Messages
Suggested actions
Equilibrate and restart module.
Please contact your Agilent service
representative.
Suggested actions
Refill solvent bottle.
Check solvent counter setting in user interface.