Degasser's pressure limit violation

Error ID: 29220
Pressure too far above the limit.
Probable cause
1
2
3

Seal wash pump was missing when tried to turn on

Error ID: 29223
The seal wash pump has not been detected (while being configured or
detected before)
Probable cause
1
2
3
Agilent 1290 Infinity II Flexible Pump User Manual
Leak in degasser chamber or degasser
tubing.
Defect vacuum pump.
Degasser chamber empty or connected to
air.
Defect cable connection to seal wash pump.
Defect seal wash pump motor.
Defective main board.
Error Information
Pump Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Block unused degasser channels.
Suggested actions
Check cable connection.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.