Agilent Technologies IDP-10 Instruction Manual

Agilent Technologies IDP-10 Instruction Manual

Dry scroll vacuum pump
Hide thumbs Also See for IDP-10:

Advertisement

Quick Links

Vacuum Products Division
IDP-10
INSTRUCTION MANUAL
Dry Scroll
Vacuum Pump
Manual No. X3810-90000
Revision B
January 2017

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies IDP-10

  • Page 1 Vacuum Products Division IDP-10 INSTRUCTION MANUAL Dry Scroll Vacuum Pump Manual No. X3810-90000 Revision B January 2017...
  • Page 2 IDP-10 Dry Scroll Vacuum Pump Slo-Blo and Littelfuse are registered trademarks of Littelfuse, Inc. Copyright 2017 Agilent, Inc.
  • Page 3: Warranty

    IDP-10 Dry Scroll Vacuum Pump Warranty Warranty Replacement Products manufactured by Seller are warranted against defects in materials and workmanship for twelve (12) and Adjustment months from date of shipment thereof to Customer, and Seller's liability under valid warranty claims is limited, at...
  • Page 4 IDP-10 Dry Scroll Vacuum Pump This page intentionally left blank.
  • Page 5: Table Of Contents

    Interface de l'utilisateur ....10 IDP-10 Power Cord Selection ... 18 Opération ......11 Typical IDP-10 Maximum Starting Fonctions de protection .
  • Page 6 IDP-10 Dry Scroll Vacuum Pump This page intentionally left blank.
  • Page 7: Instructions For Use

    The IDP-10 is a hermetic, dry scroll vacuum pump. This pump is suitable for pumping air or inert gases. The pump Total weight of the packing, IDP-10 pump included, is is not intended to pump corrosive, explosive, or particu- approximately 28.1 kg (62 lbs).
  • Page 8: Emc Warnings

    The IDP-10 was tested with an I/O cable (for remote con- NOTE The equipment has been tested and...
  • Page 9: Installation

    IDP-10 Dry Scroll Vacuum Pump Installation The IDP-10 is designed and certified for operation The IDP-10 employs an inverter driven 3-phase motor so indoors (installation category II) in relatively clean, dry that pump speed is independent of AC mains frequency.
  • Page 10: User Interface

    IDP-10 Dry Scroll Vacuum Pump User Interface Description of Pin Functions ❑ Pin#1: This pin can be used to control the speed of the There is a slide switch, used for local manual operation of pump by applying an analog voltage between 1 and the pump, and a DSUB9 type connector with male pin either of the signal ground pins.
  • Page 11: Operation

    IDP-10. A cable with both after a time delay to reduce power dissipation. After foil and braid shielding is recom-...
  • Page 12: Maintenance

    IDP-10 Dry Scroll Vacuum Pump Maintenance Disposal Meaning of the WEEE logo found in labels Personnel responsible for pump operation and mainte- The following symbol is applied in accordance with the nance must be well-trained and aware of the accident EC WEEE (Waste Electrical and Electronic Equipment) prevention rules.
  • Page 13: Environmental Conditions, Compliance And Utility

    ❑ Changes or modifications not expressly approved by NOTE Pollution level describes the degree to Agilent Technologies could void the user's authority to which a solid, liquid, or gas which dete- operate the equipment. riorates dielectric strength is adhering. 2 applies to a normal indoor atmosphere.
  • Page 14: Instructions D'utilisation

    L'IDP-10 est une pompe à vide à spirales à sec her- Le poids total de l’emballage, pompe IDP-10 incluse, est d’environ 28,1 kg. métique. Cette pompe est adaptée au pompage de l'air ou AVERTISSEMENT La pompe IDP-10 elle-même pèse des gaz inertes.
  • Page 15: Avertissements Cem

    L’IDP-10 a été testée avec un câble E/S (pour la com- REMARQUE L’équipement a été testé et déclaré con- mande distante), en employant une perle ferrite à pince, forme aux limites d’un dispositif...
  • Page 16: Installation

    Installation Interface de l'utilisateur Il y a un interrupteur coulissant, utilisé pour l'opération L'IDP-10 est conçue et certifiée pour une opération à manuelle locale de la pompe et un connecteur de type l'intérieur (catégorie d'installation 1) dans des environne- DSUB9 avec les contacts de broche mâles pour la com- ments relativement propres, secs (degré...
  • Page 17: Opération

    Si la pompe dispose de la vanne d'admission en option, utilisé lors de l'utilisation d'un potentiomètre pour le elle s'ouvre dans environ 30 secondes et l'IDP-10 com- contrôle de la vitesse. Lorsque vous raccordez le haut mence à pomper. Si la pompe n'est pas présente, le pom- d'un potentiomètre à...
  • Page 18: Fonctions De Protection

    IDP-10 Dry Scroll Vacuum Pump Fonctions de protection AVERTISSEMENT La mort peut résulter d'un contact avec des tensions élevées. Faites tou- L'IDP-10 a plusieurs fonctions de protection: jours extrêmement attention et ❑ Contrôle de la surcharge : lorsque la pompe est sous respectez les règlementations de...
  • Page 19: Technical Information

    IDP-10 Dry Scroll Vacuum Pump Technical Information Table 5 Specifications Model IDP-10 Dry Scroll Single Hermetic Vacuum Pump Interface dimensions H=297 mm x W=260 mm x L= 420mm Peak pumping speed Default factory setting: 63.3 Hz: 1900 RPM, 170 L/min (10.2 m...
  • Page 20: Outline Drawing With Dimensions

    IDP-10 Dry Scroll Vacuum Pump Figure 4 Outline Drawing with Dimensions Figure 5 Pumping Speed Curves...
  • Page 21: Unpacking And Inspection

    Unpacking and Inspection 1. Orient the shipping container with This End Up on top. 2. Open the box and carefully lift the IDP-10 out of the box. 3. Save the carton and all packing materials. 4. Inspect the pump for damage. If there is shipping...
  • Page 22: Installation

    IDP-10 Dry Scroll Vacuum Pump Installation Startup 1. Check that the inlet screen is installed before begin- Safety ning operation. Do not remove or modify any safety or insulating equip- WARNING Do not insert a finger or any foreign ment from the pump. To do so may create a serious safety object in the path of the fan;...
  • Page 23: Electrical Connections

    Table 6. Do not modify the plug provided; if it does not fit the outlet, have the proper Table 6 IDP-10 Power Cord Selection outlet installed by a qualified electrician. Country Power Cord Specification...
  • Page 24: Start Current

    IDP-10 Dry Scroll Vacuum Pump Start Current Table 7 lists the typical IDP-10 maximum starting current and duration. Table 7 Typical IDP-10 Maximum Starting Current (A) Frequency High Voltage Range 100 V 120 V 127 V 200 V 230 V 240 V...
  • Page 25: Mechanical Connections

    Isolation Valve The IDP-10 has an optional integral automatic isolation valve. If the IDP-10 is equipped with this valve, an inter- nal timer opens the valve 30 seconds after the pump is switched ON. If power is lost or the pump is switched Gas ballast port 1 is used for moderate water vapor loads, OFF, the isolation valve will immediately close.
  • Page 26: Operation

    IDP-10 Dry Scroll Vacuum Pump Operation Cleaning the Pump Unlike conventional oil-sealed pumps, Agilent dry scroll pumps do not contain fluid for the cleansing of accumu- lated dust and debris. Run the pump periodically at atmo- sphere for a minute or two to flush it out. Until...
  • Page 27: Troubleshooting

    Replacement The IDP-10 is equipped with an auto-reset thermal motor protector. This protector will automatically shut down the pump when it detects an overload condition, and will automatically restart the pump when the motor has cooled to within an acceptable tem-...
  • Page 28: Maintenance

    Table 9 Accessories and Replacement Kits Cleaning Part Description Contents Exterior Number The exterior surfaces of the IDP-10 may be cleaned with a X3807-67002 Front Cowling Replacement front cloth slightly moistened with water. If necessary alcohol Assembly cowling material to or a mild detergent may be used instead.
  • Page 29: Exploded View Of Pump Body

    IDP-10 Dry Scroll Vacuum Pump 3. Using a crescent wrench, remove the exhaust adapter To install the new seals and O-rings: (item 2). 1. Unpack the Tip Seal. 4. Remove the front cowling and hang out of the way 2. Install the closed loop portion of the Tip Seal onto the (item 3).
  • Page 30 IDP-10 Dry Scroll Vacuum Pump 1. Front Cowling Bolts; M8 (3) 2. Exhaust Fitting (NW16) 3. Front Cowling Assembly 4. Frame Bolts; M5 (4) 5. Scroll Housing 6. Main O-ring; Parker 2-265 Viton Tip Seal for Scroll Housing 8. Tip Seal for Orbiting Scroll 9.
  • Page 31: Request For Return Health And Safety Certification

    Please follow these instructions whenever one of our products needs to be returned. Complete the attached R R equest for Return form and send it to Agilent Technologies (see below), taking particular care to include the completed H H ealth and Safety declaration Section. No work can be started on your unit until we receive a completed copy of this form.
  • Page 32 Terms and conditions TERMS AND CONDITIONS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division – Products and Services Terms of Sale. Unless otherwise pre-negotiated, customer is responsible for the freight charges for the returning product. Return shipments must comply with all applicable S S hipping Regulations (IATA, DOT, etc.) and carrier requirements.
  • Page 33 IDP-10 Dry Scroll Vacuum Pump Vacuum Products Division Request for Return Form Customer information Company : Contact Name: Address: Tel: Fax: Email: Equipment Product description Agilent PartNo Agilent Serial No Original Purchasing Reference Failure description Type of process (for which the equipment was used)
  • Page 34 IDP-10 Dry Scroll Vacuum Pump This page intentionally left blank.
  • Page 35 IDP-10 Dry Scroll Vacuum Pump This page intentionally left blank.
  • Page 36 Asia PACIFIC RIM: A ilent Technolo ies Sales Sdn Bhd please visit our website for individual office information http://www.agilent.com/chem/vacuum Worldwide Web Site, Catalog and Order On- © Agilent Technologies, Inc. 2016 line: www.agilent.com/chem/vacuum vps-customerservice@agilent.com Representatives in most countries...

Table of Contents