IDP-10 Dry Scroll Vacuum Pump Warranty Warranty Replacement Products manufactured by Seller are warranted against defects in materials and workmanship for twelve (12) and Adjustment months from date of shipment thereof to Customer, and Seller's liability under valid warranty claims is limited, at...
Page 4
IDP-10 Dry Scroll Vacuum Pump This page intentionally left blank.
The IDP-10 is a hermetic, dry scroll vacuum pump. This pump is suitable for pumping air or inert gases. The pump Total weight of the packing, IDP-10 pump included, is is not intended to pump corrosive, explosive, or particu- approximately 28.1 kg (62 lbs).
IDP-10 Dry Scroll Vacuum Pump Installation The IDP-10 is designed and certified for operation The IDP-10 employs an inverter driven 3-phase motor so indoors (installation category II) in relatively clean, dry that pump speed is independent of AC mains frequency.
IDP-10 Dry Scroll Vacuum Pump User Interface Description of Pin Functions ❑ Pin#1: This pin can be used to control the speed of the There is a slide switch, used for local manual operation of pump by applying an analog voltage between 1 and the pump, and a DSUB9 type connector with male pin either of the signal ground pins.
IDP-10 Dry Scroll Vacuum Pump Maintenance Disposal Meaning of the WEEE logo found in labels Personnel responsible for pump operation and mainte- The following symbol is applied in accordance with the nance must be well-trained and aware of the accident EC WEEE (Waste Electrical and Electronic Equipment) prevention rules.
❑ Changes or modifications not expressly approved by NOTE Pollution level describes the degree to Agilent Technologies could void the user's authority to which a solid, liquid, or gas which dete- operate the equipment. riorates dielectric strength is adhering. 2 applies to a normal indoor atmosphere.
L'IDP-10 est une pompe à vide à spirales à sec her- Le poids total de l’emballage, pompe IDP-10 incluse, est d’environ 28,1 kg. métique. Cette pompe est adaptée au pompage de l'air ou AVERTISSEMENT La pompe IDP-10 elle-même pèse des gaz inertes.
L’IDP-10 a été testée avec un câble E/S (pour la com- REMARQUE L’équipement a été testé et déclaré con- mande distante), en employant une perle ferrite à pince, forme aux limites d’un dispositif...
Installation Interface de l'utilisateur Il y a un interrupteur coulissant, utilisé pour l'opération L'IDP-10 est conçue et certifiée pour une opération à manuelle locale de la pompe et un connecteur de type l'intérieur (catégorie d'installation 1) dans des environne- DSUB9 avec les contacts de broche mâles pour la com- ments relativement propres, secs (degré...
Si la pompe dispose de la vanne d'admission en option, utilisé lors de l'utilisation d'un potentiomètre pour le elle s'ouvre dans environ 30 secondes et l'IDP-10 com- contrôle de la vitesse. Lorsque vous raccordez le haut mence à pomper. Si la pompe n'est pas présente, le pom- d'un potentiomètre à...
IDP-10 Dry Scroll Vacuum Pump Fonctions de protection AVERTISSEMENT La mort peut résulter d'un contact avec des tensions élevées. Faites tou- L'IDP-10 a plusieurs fonctions de protection: jours extrêmement attention et ❑ Contrôle de la surcharge : lorsque la pompe est sous respectez les règlementations de...
Unpacking and Inspection 1. Orient the shipping container with This End Up on top. 2. Open the box and carefully lift the IDP-10 out of the box. 3. Save the carton and all packing materials. 4. Inspect the pump for damage. If there is shipping...
IDP-10 Dry Scroll Vacuum Pump Installation Startup 1. Check that the inlet screen is installed before begin- Safety ning operation. Do not remove or modify any safety or insulating equip- WARNING Do not insert a finger or any foreign ment from the pump. To do so may create a serious safety object in the path of the fan;...
Table 6. Do not modify the plug provided; if it does not fit the outlet, have the proper Table 6 IDP-10 Power Cord Selection outlet installed by a qualified electrician. Country Power Cord Specification...
IDP-10 Dry Scroll Vacuum Pump Start Current Table 7 lists the typical IDP-10 maximum starting current and duration. Table 7 Typical IDP-10 Maximum Starting Current (A) Frequency High Voltage Range 100 V 120 V 127 V 200 V 230 V 240 V...
Isolation Valve The IDP-10 has an optional integral automatic isolation valve. If the IDP-10 is equipped with this valve, an inter- nal timer opens the valve 30 seconds after the pump is switched ON. If power is lost or the pump is switched Gas ballast port 1 is used for moderate water vapor loads, OFF, the isolation valve will immediately close.
IDP-10 Dry Scroll Vacuum Pump Operation Cleaning the Pump Unlike conventional oil-sealed pumps, Agilent dry scroll pumps do not contain fluid for the cleansing of accumu- lated dust and debris. Run the pump periodically at atmo- sphere for a minute or two to flush it out. Until...
Replacement The IDP-10 is equipped with an auto-reset thermal motor protector. This protector will automatically shut down the pump when it detects an overload condition, and will automatically restart the pump when the motor has cooled to within an acceptable tem-...
Table 9 Accessories and Replacement Kits Cleaning Part Description Contents Exterior Number The exterior surfaces of the IDP-10 may be cleaned with a X3807-67002 Front Cowling Replacement front cloth slightly moistened with water. If necessary alcohol Assembly cowling material to or a mild detergent may be used instead.
IDP-10 Dry Scroll Vacuum Pump 3. Using a crescent wrench, remove the exhaust adapter To install the new seals and O-rings: (item 2). 1. Unpack the Tip Seal. 4. Remove the front cowling and hang out of the way 2. Install the closed loop portion of the Tip Seal onto the (item 3).
Page 30
IDP-10 Dry Scroll Vacuum Pump 1. Front Cowling Bolts; M8 (3) 2. Exhaust Fitting (NW16) 3. Front Cowling Assembly 4. Frame Bolts; M5 (4) 5. Scroll Housing 6. Main O-ring; Parker 2-265 Viton Tip Seal for Scroll Housing 8. Tip Seal for Orbiting Scroll 9.
Please follow these instructions whenever one of our products needs to be returned. Complete the attached R R equest for Return form and send it to Agilent Technologies (see below), taking particular care to include the completed H H ealth and Safety declaration Section. No work can be started on your unit until we receive a completed copy of this form.
Page 32
Terms and conditions TERMS AND CONDITIONS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division – Products and Services Terms of Sale. Unless otherwise pre-negotiated, customer is responsible for the freight charges for the returning product. Return shipments must comply with all applicable S S hipping Regulations (IATA, DOT, etc.) and carrier requirements.
Page 33
IDP-10 Dry Scroll Vacuum Pump Vacuum Products Division Request for Return Form Customer information Company : Contact Name: Address: Tel: Fax: Email: Equipment Product description Agilent PartNo Agilent Serial No Original Purchasing Reference Failure description Type of process (for which the equipment was used)