Prepare the Pump Module for Transport

When
If the module shall be transported or shipped.
Parts required
p/n
9301-0411
9301-1337
G7104-44000
Mechanical damage
C A U T I O N
1
Flush all solvent channels with isopropanol.
2
Remove solvent inlet tubes from solvent reservoirs and
tubing clips at other modules.
3
Remove tubings between the seal wash function and
solvent bottle/waste.
4
Turn off the pump.
5
Remove cable and capillary connections to other
modules.
6
Remove the waste tube.
7
Remove the module from the stack.
Agilent 1290 Infinity II Flexible Pump User Manual
Description
Syringe; Plastic
Syringe adapter
Transport protection foam
For shipping the module, insert the Protective Foam to protected the module from
mechanical damage.
Be careful not to damage tubing or capillary connections while inserting the module
in the Protective Foam.

Prepare the Pump Module for Transport

8
Disconnect the degasser outlet tubings at the MCGV one
after another. Use a syringe for removing liquid from the
degasser and the solvent tubings.