Setting the 8-bit Configuration Switch

Setting the 8-bit Configuration Switch

The 8-bit configuration switch is located at the rear of the module. Switch
settings provide configuration parameters for LAN, serial communication
protocol and instrument specific initialization procedures.
All modules with on-board LAN:
For normal operation use the default (best) settings.
NOTE
Figure 25
To perform any LAN configuration, SW1 and SW2 must be set to OFF. For details
NOTE
on the LAN settings/configuration refer to chapter LAN Configuration.
1290 Infinity II High-Speed Pumps User Manual
Default is ALL switches DOWN (best settings).
19200 baud, 8 data bit / 1 stop bit with no parity for RS-232
For specific LAN modes switches 3-8 must be set as required.
For boot/test modes switches 1+2 must be UP plus required mode.
Location of Configuration Switch (example shows a G4212A DAD)