Timeout: Wait For Flow; Timeout: Wait For Pressure - Agilent Technologies Infinity II 1290 User Manual

Flexible pump
Hide thumbs Also See for Infinity II 1290:
Table of Contents

Advertisement

Timeout: Wait for Flow

Error ID: 29183
A target condition (flow rate) has been sent to the instrument which
should have been reached within an expected time frame but didn't.
Either the limit, time frame or the current value of the variable has been
modified later directly or indirectly.
Probable cause
1

Timeout: Wait for Pressure

Error ID: 29185
A target condition (pressure) has been sent to the instrument which
should have been reached within an expected time frame but didn't.
Either the limit, time frame or the current value of the variable has been
modified later directly or indirectly.
Probable cause
1
2
Agilent 1290 Infinity II Flexible Pump User Manual
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Leak
Error Information
Pump Error Messages
Suggested actions
Verify control software, macros, manual
commands.
Suggested actions
Verify control software, macros, manual
commands.
Run system pressure test for identifying and
localizing the leak. Tighten leak.
6
75