Firmware updates can be done using the following tools (latest version
should be used):
(*)
• Agilent Lab Advisor software with files on the hard disk
(*)
• Firmware Update Tool with local files on the hard disk
• Instant Pilot (G4208A) with files on a USB Flash Disk
(*)
Required tools, firmware and documentation are available from the
Agilent web:
http://www.chem.agilent.com/_layouts/agilent/downloadFirmware.aspx?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
PPPP is the product number, for example, 1315B for the G1315B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C
DAD,
VVV is the revision number, for example 650 is revision 6.50,
XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware
in chapter "Maintenance" or use the documentation provided with the
Firmware Update Tools.
Update of main system can be done in the resident system only. Update of the resident
N O T E
system can be done in the main system only.
Main and resident firmware must be from the same set.