Reading of multi purpose valve tag failed
Error ID: 29240
Reading the multi purpose valve tag failed.
• Parameter: None
Probable cause
1
2
3
Pump drive encoder rollover
Error ID: 29232
Invalid pump drive encoder signals have been detected.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
Agilent 1290 Infinity II Flexible Pump User Manual
Reading of multi purpose valve tag failed.
Multi purpose valve head tag defect or
empty.
Multi purpose valve tag reader is defect.
Pump drive encoder is defect.
Error Information
Pump Error Messages
Suggested actions
Check cable connection.
Replace multi purpose valve head.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
6