Agilent Technologies Infinity II 1290 User Manual page 196

Flexible pump
Hide thumbs Also See for Infinity II 1290:
Table of Contents

Advertisement

11
LAN Configuration
Initialization Mode Selection
Since the default IP address is a so- called local address, it will not be
routed by any network device. Thus, the PC and the module must reside
in the same subnet.
The user may open a Telnet session using the default IP address and
change the parameters stored in the non- volatile memory of the module.
He may then close the session, select the initialization mode Using Stored,
power- on again and establish the TCP/IP connection using the new
parameters.
When the module is wired to the PC directly (e.g. using a cross- over cable
or a local hub), separated from the local area network, the user may
simply keep the default parameters to establish the TCP/IP connection.
In the Using Default mode, the parameters stored in the memory of the module are not
N O T E
cleared automatically. If not changed by the user, they are still available, when switching
back to the mode Using Stored.
196
Agilent 1290 Infinity II Flexible Pump User Manual