Remove the Jet Weaver

Tools required
p/n
8710-0510
8710-0899
Parts required
#
2
Preparations
• Select Do not use mixer in ChemStation.
• Switch off the pump at the main power switch.
1
Remove capillary connections from the Jet Weaver to the
Multi Purpose Valve.
Agilent 1290 Infinity II Flexible Pump User Manual
Description
Wrench open 1/4 — 5/16 inch
Pozidriv screwdriver
p/n
Description
0100-1259
Plastic fittings
2
Open the screw which fixes the Jet Weaver to the front
panel.
Maintenance

Remove the Jet Weaver