Error Information

General Error Messages

General Error Messages

General error messages are generic to all Agilent series HPLC modules
and may show up on other modules as well.

Timeout

Error ID: 0062
The timeout threshold was exceeded.
Probable cause
1
2

Shutdown

Error ID: 0063
An external instrument has generated a shutdown signal on the remote
line.
The module continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates
the error message.
54
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Suggested actions
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.