Hardware Information

Setting the 8-bit Configuration Switch (without On-board) LAN

Setting the 8-bit Configuration Switch (without On-board) LAN

The 8- bit configuration switch is located at the rear of the module.
This module does not have its own on- board LAN interface. It can be
controlled through the LAN interface of another module, and a CAN
connection to that module.
Figure 38
All modules without on- board LAN:
• default should be ALL DIPS DOWN (= best settings)
• DIP 1 DOWN and DIP 2 UP allows special RS- 232 settings
• for boot/test modes DIPS 1+2 must be UP plus required mode
For normal operation use the default (best) settings.
N O T E
Switch settings provide configuration parameters for serial communication
protocol and instrument specific initialization procedures.
With the introduction of the Agilent 1260 Infinity, all GPIB interfaces have been removed.
N O T E
The preferred communication is LAN.
228
Configuration switch (settings depend on configured mode)
Bootp mode for LAN and
19200 baud, 8 data bit / 1 stop bit with no parity for RS- 232
1260 Infinity Binary Pump User Manual