Setting the 8-bit Configuration Switch

Special Settings

The special settings are required for specific actions (normally in a service case).
The tables include both settings for modules – with on-board LAN and without on-board
N O T E
LAN. They are identified as LAN and no LAN.
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up
again, the instrument firmware stays in the resident mode. It is not
operable as a module. It only uses basic functions of the operating system
for example, for communication. In this mode the main firmware can be
loaded (using update utilities).
Table 8
Boot Resident Settings (On-board LAN)
Mode Select
TEST/BOOT
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode
with default parameter settings.
Loss of data
C A U T I O N
Forced cold start erases all methods and data stored in the non-volatile memory. Exceptions
are calibration settings, diagnosis and repair log books which will not be erased.
If you use the following switch settings and power the instrument up
again, a forced cold start has been completed.
Table 9
Forced Cold Start Settings (On-board LAN)
Mode Select
TEST/BOOT
186
SW1
SW2
SW3
1
1
Save your methods and data before executing a forced cold start.
SW1
SW2
SW3
1
1
SW4
SW5
1
0
0
SW4
SW5
0
0
0
Agilent 1290 Infinity II Flexible Pump User Manual
SW6
SW7
SW8
0
0
SW6
SW7
SW8
0
0
0
1