Writing the multi purpose valve tag failed

Error ID: 29241
Writing the multi purpose valve tag failed.
• Parameter: None
Probable cause
1
2

Unknown multi purpose valve type

Error ID: 29242
The type information of the multi purpose valve is invalid.
• Parameter: None
Probable cause
1
2

Pump drive encoder error

Error ID: 29211
The pump drive encoder has generated no signal.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
Agilent 1290 Infinity II Flexible Pump User Manual
Multi purpose valve head tag defect.
Multi purpose valve tag head reader is
defect.
Wrong valve head installed.
Valve head has invalid RFID tag content.
Pump drive encoder is defect.
Error Information
Pump Error Messages
Suggested actions
Replace multi purpose valve head.
Please contact your Agilent service
representative.
Suggested actions
Check or replace multi purpose valve head.
Check or replace multi purpose valve head.
Suggested actions
Please contact your Agilent service
representative.