Deliver Underrun; Defect Connection Between Main Board And Pump Drive Encoder - Agilent Technologies Infinity II 1290 User Manual

Flexible pump
Hide thumbs Also See for Infinity II 1290:
Table of Contents

Advertisement

Deliver underrun

Error ID: 29233
Internal error.
• Parameter: None
Probable cause
1
2

Defect connection between main board and pump drive encoder

Error ID: 29208
Defect connection between main board and pump drive encoder.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2
Agilent 1290 Infinity II Flexible Pump User Manual
Internal error.
Firmware issue
Defect connection between main board and
pump drive encoder.
Defect encoder.
Error Information
Pump Error Messages
Suggested actions
Please contact your Agilent service
representative.
Use a minimum firmware revision of B.06.55
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
6