Drive timeout

Error ID: 29204
Movement of drive during initialization is blocked mechanically.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2

Overcurrent of pump drive

Error ID: 29202
The current consumption of the pump drive is too high.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2
66
Blockage in flow path
Blockage of pump drive Drive motor defect.
Blockage of system before pressure sensor.
Drive motor defect.
Suggested actions
Remove capillary connection to system, check
outlet filter, check valves, check pump head.
Please contact your Agilent service
representative.
Suggested actions
Check for blockage of e.g. outlet valve filter frit,
Multi Purpose Valve, heat exchanger.
Please contact your Agilent service
representative.
Agilent 1290 Infinity II Flexible Pump User Manual