Use two 5/16 wrenches for opening the inline filter.
5
Put the inline filter to the clamp and install its capillaries.
The integrated capillary is connected to port 5 of the
Multi Purpose Valve. The removable capillary is
connected to port 8.
Agilent 1290 Infinity II Flexible Pump User Manual
Replace Parts of the Inline Filter
4
Replace the filter frit and reassemble the inline filter.