Replace Parts Of The Multi Purpose Valve - Agilent Technologies Infinity II 1290 User Manual

Flexible pump
Hide thumbs Also See for Infinity II 1290:
Table of Contents

Advertisement

Replace Parts of the Multi Purpose Valve

Tools required
p/n
8710-2394
Parts required
p/n
1535-4045
5068-0202
5068-0120
5068-0001
1535-4857
Preparations
• Remove capillary connections from ports 1, 3 and 6.
• Remove the clamp with the inline filter.
1 Use the 9/64 inch hex key for opening the valve head.
2 Replace parts as required.
3 Reassemble the valve head and mount it to the valve drive.
Agilent 1290 Infinity II Flexible Pump User Manual
Description
9/64 inch hex key
Description
Bearing ring
Rotor seal, Multi Purpose Valve, 1300 bar
Stator ring
Stator head
Stator screws, 10/Pk

Replace Parts of the Multi Purpose Valve

7
Maintenance
135

Advertisement

Table of Contents
loading

This manual is also suitable for:

G7104a

Table of Contents