4
Installation
Hardware Installation
1 Mount the clamp guide on the right side of the Infinity II Pump: Markings in the
2 Mount the valve heads on the G1170A external valve drives.
3 Clamp the first external valve drive with the MHC valve on top.
4 Attach the adapter-profile on each of the other external valve drives and
5 Mount the leak tray with sensor underneath the lowest external valve drive.
6 Install the Pressure release kit, see
2D-LC User Guide
form of round dips are on the body housing. Make a small hole with a peaked
screw driver and tighten the clamp guide with the 3 self-cutting tapping
screws.
mount them according to the positions shown in
page 85.
Figure 25
"Installing the Pressure Release Kit"
on page 61.
on