Agilent Technologies 1290 Infinity II 2D-LC User Manual page 221

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

8
Run the System
Run the Experiment
Table 29
Recommended conditions in 2D (HPLC) for ASM MHC 2D-LC
Parameter
Value
Sampling Table
Start 0.2 min Time based, 2.5 min Peak Based, 4.50 min
2D Gradient
Analysis 1.50 min Equilibration 0.50 min Cycle time 2.12 with ASM ON and ASM Factor 3
Analytical gradient - Shifted Gradient Shift 1D:
3 % B 0.00 min
3 % B 0.37 min
10 % B 0.38 min
60 % B 1.62
2
D Detector (DAD)
Diode-array
254 nm, Bandwidth 4 nm
Reference Wavelength 360 nm
Reference Bandwidth
100 nm
Peak width
80 Hz
Stop time
As pump/No limit
2D-LC User Guide
30 % B 6 min
221