8
Run the System
Run the Experiment
Table 34
Parameter
Average Scans
Detection Window
(ppm)
Min Height
Stop time
To avoid problems in the LC/MS due to the high flow rate the effluent from the second dimension
column should be split. The recommended split ratio is 1:2
Active ASM for comprehensive applications is not recommended as the wear of
NOTE
the valve increases dramatically due to the many switching cycles.
1 Load method ASM Comprehensive Checkout from the 2D-LC data media and
2 Run the method with 2D-LC starter sample, 1 x 2 mL (5190-6895), 1:10 (for
3 If necessary, subsequently edit or optimize the method.
2D-LC User Guide
Recommended conditions in 2D (LC-MS)
Value
AutoRecalibration
1
100 ppm
1000 counts
Reference Masses
Positive
121.05087300
922.00979800
Chromatograms
Chrom Type Label Offset Y-Range
TIC TIC 1510000000
TIC TIC 1510000000
As pump/No limit
modify the settings for your configuration.
only UV Checkout), 1:100 (for LCMS Checkout), or 1:1000 (for LCMS
Checkout) diluted with Methanol/Water (20/80; v/v) with 0.1 % formic acid.