Agilent Technologies 1290 Infinity II 2D-LC User Manual page 293

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

10
Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
To upgrade/downgrade the module's firmware carry out the following steps:
1 Download the required module firmware, the latest Lab Advisor software and
2 For loading the firmware into the module
Do not interrupt the power supply of the device and the PC during this procedure.
NOTE
2D-LC User Guide
the documentation from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
a Select the folder on the hard drive where the Firmware package is stored.
b Connect the Lab Advisor Software to your 2D-LC instrument.
Figure 153 Firmware Update
c Press the Lock button.
The system is locked.
d Select the required firmware version for the Resident and Main Firmware.
e To update the firmware of the instrument, press the Update button.
This will require some time.
293