Agilent Technologies 1290 Infinity II 2D-LC User Manual page 213

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

8
Run the System
Run the Experiment
Table 23
Parameter
Min Height
Stop time
1 Load method High Resolution Checkout from the 2D-LC data media and
2 Run the method with 2D-LC starter sample, 1 x 2 mL (5190-6895), 1:10 (for
3 If necessary, subsequently edit or optimize the method.
2D-LC User Guide
Recommended conditions in 2D (LC-MS)
Value
1000 counts
Reference Masses
Positive
121.05087300
922.00979800
Chromatograms
Chrom Type Label Offset Y-Range
TIC TIC 1510000000
TIC TIC 1510000000
As pump/No limit
modify the settings for your multiple heart cutting configuration.
only UV Checkout), 1:100 (for LCMS Checkout), or 1:1000 (for LCMS
Checkout) diluted with Methanol/Water (20/80; v/v) with 0.1 % formic acid.
213

Advertisement

Table of Contents
loading

This manual is also suitable for:

1290 infinity ii lc

Table of Contents