8
Run the System
Run the Experiment
Table 33
Recommended conditions in 2D (HPLC) for ASM comprehensive 2D-LC
Parameter
Value
2D-LC Valve
MHC with 40 µl sample, Transfer Capillary, ASM Factor No
2
D Column Compartment (MCT)
Column
RRHD Bonus-RP, 2.1x 50 mm, 1.8 µm, 1200 bar (857768-901)
Column temperature
40 °C
Stop time
As pump/No limit
2
D Pump
Comprehensive
Figure 112 Comprehensive
Mobile Phase A
Water + 0.2 % formic acid
Mobile Phase B
Methanol
Flow Rate
2.5 mL/min
Stop time
40 min (will not automatically prolonged, if peaks in 2D are not work off)
Post time
6 min
2D gradient
Analysis 0.2 min Equilbration 0.1 minASM Off
Analytical gradient - shifted gradient shift 1D
25 % B 0.00 min
50 % B 40 min
50 % B 0.20 min
75 % B 40 min
2
D Detector (DAD)
Diode-array
254 nm, Bandwidth 4 nm
Reference Wavelength 360 nm
Reference Bandwidth
100 nm
Peak width
80 Hz
Stop time
As pump/No limit
2D-LC User Guide
25 % B 5 min
50 % B 5 min
226