Method Parameters
Set the 2D-LC Method parameters
[OPTIONAL]
3 Set the maximum and minimum Pressure Limits for the pump.
The default settings are recommended. Change these settings only for important
NOTE
and valid reasons.
For further details, especially the pressure limits, see the user manual of your
NOTE
pump.
2D-LC User Guide
•
Max is the maximum pressure limit at which the pump will switch itself off.
This maximum pressure limit protects the system against overpressure.
•
Min is the minimum pressure limit at which the pump will switch itself off.
This situation can occur, when a solvent reservoir is empty.
The minimum pressure limit protects the system from damage caused by
pumping air.
153