Run the System
Configure the 2D-LC System
Configure 2D-LC Hardware
Focus on the 2D-LC Valves and the capillary connection.
1 To find out the correct plumbing of the 2D-LC valve ports, see
Table 12
2D-LC mode
Standard heart-cutting
MHC or HiRes
Comprehensive
40 µL sample loops are part of the default setup in the methods of the data
NOTE
media.
The 2D-LC valve with one single loop setup is only used for special applications,
NOTE
for example the Bio ProA-Sec Kit.
For more information, see the bio application documentation.
Methods for preparation and checkout runs of recommended system
NOTE
configurations are available on the Agilent 2D-LC Software data media.
Not all possible configurations can be shown here. Therefore, adapt these
methods for other configurations and modules if necessary.
2D-LC User Guide
2D-LC Valve, Standard (G4236A)"
ASM (G4243A)"
on page 70, or the 2D-LC online help.
The recommended plumbing of the 2D-LC valve differs between 2D-LC setup
with single loops versus 2D-LC setup with Multi Heart Cutting (MHC) Valves
and concurrent versus countercurrent mode.
Hardware setups for 2D-LC modi
Hardware setup
on page 67,
2D-LC Valve with one single loop
2D-LC Valve with two single loops
2D-LC Valve with two MHC valves (each with six Sample Loops)
2D-LC ASM Valve with two MHC valves (each with six Sample
Loops)
2D-LC Valve with two MHC valves (each with six Sample Loops)
2D-LC ASM Valve with two MHC valves (each with six Sample
Loops)
2D-LC Valve with two single loops and 2D-LC Valve with two MHC
valves (each with six Sample Loops)
2D-LC ASM Valve with two MHC valves (each with six Sample
Loops)
"Connecting the
"Connecting the 2D-LC Valve,