8
Run the System
Run the Experiment
Table 26
Recommended conditions in 2D (HPLC) for comprehensive 2D-LC
Parameter
Value
2D-LC Valve
MHC with 40 µl sample, Transfer Capillary, ASM Factor No
2
D Column Compartment (MCT)
Column
RRHD Eclipse Plus C18, 3.0x 50 mm, 1.8 µm (959757-302)
Column temperature
40 °C
Stop time
As pump/No limit
2
D Pump
Comprehensive
Mobile Phase A
Water + 0.2 % formic acid
Mobile Phase B
Acetonitrile
Flow Rate
2.5 mL/min
Idle flow
not used
Stop time
ca. 43 min (will not automatically prolonged, if peaks in 2D are not work off)
Post time
6 min
Sampling Table
Start 5 min, Stop at 40 min
2D Gradient:
Analysis 0.2 min, Equilibration 0.15 min
Analytical gradient - Shifted Gradient Shift 1D:
25 % B 0.00 min
50 % B 0.2 min
2
D Detector (DAD)
Diode-array
254 nm, Bandwidth 4 nm
Reference Wavelength 360 nm
Reference Bandwidth
100 nm
Peak width
80 Hz
Stop time
As pump/No limit
2D-LC User Guide
25 % B 5 min
50 % B 40 min
50 % B 5 min
75 % B40 min
216