Materials required:
•
Cotton swab
•
Dental mirror
•
EZ KLEEN
•
Flashlight
•
Pipette, 2-mm ID, disposable
Time: 2 minutes
Analyzer mode: Ready to Run
To clean the perox chamber
1. Remove the perox chamber cap (
2. Dip a clean, cotton-tipped swab into EZ KLEEN and wipe clean the inside of
the perox chamber.
To avoid pushing any dirt into the drain and reagent ports at the base and
back wall of the chamber, always wipe the chamber by starting at the bottom
of the back wall, using a circular up and outward motion.
3. Using a 2-mm disposable pipette, fill the chamber 1/3 full with EZ KLEEN.
4. Using the pipette, agitate the liquid in the chamber, then let it soak for one
minute.
5. Agitate the liquid again. Then, using the pipette, remove all the liquid from
the chamber.
6. Check if there are any black particles in the liquid. If yes, repeat steps 3
through 6 until the EZ KLEEN is clear when you remove it from the
chamber.
7. Using the dental mirror and flashlight, check that the chamber is clean. If it is
not clean, repeat steps 3 through 6.
8. Replace the perox chamber cap.
9. Perform a System Wash.
Cleaning the Autosampler Rinse-to-Waste Line (V45 Pathway)
Materials required
•
Beaker
•
Deionized water
5-8
CAUTION
1 PEROX CHAMBER CAP
2 PEROX CHAMBER
).
1
Troubleshooting the Analyzer