Initialization Failed - Writing Image Download
Software communication error or possible faulty Ethernet cable (from the
analyzer - J2 Workstation connection to the computer).
Corrective Action
Follow these steps to exit the ADVIA 2120/2120i software and then restart:
1
Select
2
Select
software.
If the problem persists, verify that the Ethernet connection is secure, and
replace if necessary. If the problem recurs, call Siemens Service for assistance.
Invalid / Unknown TDC Error
An unrecognized Data Manager error has occurred.
Corrective Action
Follow these steps to exit the ADVIA 2120/2120i software and then restart:
1
Select
2
Select
software.
If the problem persists, call Siemens Service for assistance.
Invalid Configuration, Mandatory Field Missing
The configuration file is incorrect. A mandatory field is missing.
Corrective Action
On the Screen Configuration window, correct the configuration file. Select
Customize menu, System Setup, Tools Modify, Screen Configuration.
Invalid Host Test Number (=0)
A test has an improper host number defined in the Test Dictionary.
Corrective Action
Verify host number in Test Dictionary, Characteristics.
Invalid Morphology Flags Host Number: xxx
The morphology flags from the analyzer cannot be managed by the Data
Manager and cannot be transmitted to the host.
7-56
Shut Down ADVIA
at the Log On/Off window.
CTRL+ALT+DELETE
Shut Down ADVIA
at the Log On/Off window.
CTRL+ALT+DELETE
and log back onto the system to restart the
and log back onto the system to restart the