Corrective Action
Define at least one QC file in the Control Dictionary window.
A
A Defined File in Prg.par Is Not Found for Backup
A file is missing that has been defined in the list of programs to be saved when
running the Program backup through the End of Day functions.
Corrective Action
Call Siemens Service for assistance.
Analyzer Connected
The computer has connected with the analyzer.
Analyzer Not Connected
The computer has not connected with the analyzer.
Possible Cause
1
The analyzer is
not on.
2
Faulty or
disconnected
Ethernet cable
(from the
analyzer - J2
Workstation
connection to
the computer).
3
Faulty analyzer
CPU board or
network
interface PC
board.
Analyzer Not Ready
The analyzer cannot start sample aspiration or hydraulic cycles.
Possible Cause
1
The SHEATH/
Status Line messages
Corrective Action
Select
at the analyzer touchpad.
On
Verify that the Ethernet cable connection is secure.
Replace if necessary.
Call Siemens Service for assistance.
Corrective Action
Replace the SHEATH/RINSE reagent.
7-7