Siemens ADVIA 2120 Operator's Manual page 213

Table of Contents

Advertisement

2
Faulty or
disconnected
Ethernet cable
(from the
analyzer - J2
Workstation
connection to
the computer)
Comparison Error MCHC/CHCM - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Comparison Error MCHC/CHCM - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Comparison Error WBCB/WBCP - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Comparison Error WBCB/WBCP - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
7-44
iv
Restart the analyzer by selecting
analyzer touchpad.
Verify that the Ethernet connection is secure. Replace if
necessary.
at the
On
Status Line messages