Reprogramming The Manual Barcode Reader; Cleaning Procedures; Backflushing The Drain Filters - Siemens ADVIA 2120 Operator's Manual

Table of Contents

Advertisement

NOTE
Achieve the final maximum lamp value in a clockwise direction only. This will
counteract any spring-bounce that might decrease the output level of the lamp
over time.
7. Once the value is maximized, tighten the securing screws (
lamp into position. Some minor drifting (less than 0.1 volts) is normal.
8. At the
If the value is at least 120 but less than 255, the perox light source indicator,
although displayed in red, is acceptable for reporting results. Results will not
be flagged. Check the gains for the perox channel and continue normal
operation.
If the value is less than 119, select
the Perox channel.

Reprogramming the Manual Barcode Reader

1. Locate the
2. Scan the
emitting a constant beeping sound, which is an indication that it is in the
programming mode.
3. Scan the
high-pitched beep to indicate that the scanning was successful.
4. Scan the
the reprogramming is finished.

Cleaning Procedures

Backflushing the Drain Filters

If a drain filter is clogged, replace it with a new filter.
If the filter is not clogged, but you want to clean it because of heavy buildup or
discoloration, use this backflush procedure.
Materials required:
Flowcell cleaning kit, PN 113-B711-01
Household bleach
Time: 10 minutes
Analyzer mode: Off
Troubleshooting the Analyzer
menu, select
Utilities
Analyzer Status
Codabar Settings
(Z7) barcode with the wand. The barcode reader should be
Start
Do not transmit Start/Stop
End
(Z7) barcode label. A continuous beeping sound indicates that
and check the lamp intensity.
Set Light Intensity
in the Opticon User's Manual.
(F0) barcode. You should hear a single
1
) to lock the
. Check the gains for
5-5