Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
Functional inspection ........... 22 5 _______ Final result/quality inspection and maintenance _____________________ 24 6 _______ Final Work Steps _______________________________________________ 25 7 _______ Changes to Previous Version_____________________________________ 26 Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 3 of 26 Medical Solutions 02.05...
General information General information Requirements The requirements described in chapter 1 of the service instructions also apply to the main- tenance described in this document. ARCOSKOP SPR2-250.831.01.03.02 Siemens AG Page 4 of 26 02.05 CS PS 24 Medical Solutions...
Systems equipped with an I.I. laser light localizer • Installation and setting instructions RXR2-120.033.03 1.2.4 For compressor, USA only • Maintenance instructions 6190800-V2-1000.02.00 • Maintenance instructions Series MDK 6 - MDK 15 Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 5 of 26 Medical Solutions 02.05 CS PS 24...
Spare parts which may be needed • Conductive rubber on the monitor trolley • Touch-up paint 34 44 403 • Spray paint 84 27 734 RH999 Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 7 of 26 Medical Solutions 02.05 CS PS 24...
When working with the laser light localizer, do not look directly into the laser beam. NOTE The eye is not in immediate danger. However, it is important not to look directly into the beam. Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 9 of 26 Medical Solutions 02.05...
NOTE The sequence for complete maintenance and inspection is de- scribed on the following pages. Each work step must be performed on an annual basis, if not oth- erwise specified. ARCOSKOP SPR2-250.831.01.03.02 Siemens AG Page 10 of 26 02.05 CS PS 24...
Maintenance should be completed according to manufacturer documentation. • The documents are located in the system folder under "maintenance". - Operating instructions Jun-Air compressor 6190800-V2-1000.02.00. - Technical documents for end-position dryer MDK 6-MDK 15. ARCOSKOP SPR2-250.831.01.03.02 Siemens AG Page 12 of 26 02.05...
Inspection of exterior and surroundings Inspection of exterior and surroundings Inspection of exterior Damage Inspect the entire system for damage, such as damage to the housing or paint. Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 13 of 26 Medical Solutions 02.05 CS PS 24...
Inspect the power outlets used for system operation for damage. • Measure the line voltage and compare it to the line voltage label on the monitor trolley. • Measure the internal line impedance. ARCOSKOP SPR2-250.831.01.03.02 Siemens AG Page 14 of 26 02.05...
Safety inspection Mechanical safety Covers • Inspect the covers of the ARCOSKOP and the monitor trolley/MSS (monitor support system) for mechanical damage and effective fastening. Cable deflectors • Inspect the cable deflectors on the monitor trolley and replace them if necessary.
Page 16
- Replace any labels that are difficult to read. • Ensure that all required ID labels are attached and in good condition. - Replace any labels that are difficult to read. Position of the labels: See ARCOSKOP operating instructions. ARCOSKOP SPR2-250.831.01.03.02 Siemens AG Page 16 of 26 02.05...
Switch on fluoroscopy for one minute. Use a wrist watch to check the fluoroscopy timer for correct operation. • All radiation indicators on the control console of the ARCOSKOP and on the monitor trolley/MSS (monitor support system) must light up during radiation. •...
Page 18
- The collimator blades must be clearly visible at the edges of the image. Dose rate • Check the dose rate. Refer to the ARCOSKOP service instructions. Area-dose product measuring device • Check the area dose product measuring device (DIAMENTOR).
Page 19
DIN VDE 0751. NOTE The equivalent leakage current test is part of the safety inspection but can only be performed when the system is closed. (Final Work Steps / p. 25). Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 19 of 26 Medical Solutions 02.05...
The fan installed in the C-arm must run audibly. Memoskop ventilation • Clean the ventilation slots on the MEMOSKOP Fast. • Ensure that the fan is functioning properly. ARCOSKOP SPR2-250.831.01.03.02 Siemens AG Page 20 of 26 02.05 CS PS 24...
Maintenance, operating value/functional inspection Operating value inspection Error memory • Read out and evaluate the error memory. Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 21 of 26 Medical Solutions 02.05 CS PS 24...
The shield of the video cable to the laser camera must not have any contact with the protective conductor of the ARCOSKOP monitor trolley. For the mea- surement briefly disconnect the connected video cable and the remote release cable.
Page 23
Then close the slot diaphragm and rotate it by +90°, +180°, -90° and -180°. In each case switch on fluoroscopy briefly and assess the position of the slot diaphragm in rela- tion to the previously displayed position of the monitor display. Slight deviations are per- mitted. Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 23 of 26 Medical Solutions 02.05...
Page 24
Final result/quality inspection and maintenance • Refit all covers. Pay attention to protective conductor connections. Image quality (IQ) quick test • Perform the ARCOSKOP image quality quick test. Maintenance • Entire system: Touch up any paint damage as needed. ARCOSKOP SPR2-250.831.01.03.02...
Page 25
Provide a copy of the certificate to the operator for filing in the system operator manual, register 9. NOTE The instructions according to ARTD-002.731.17 are to be adhered to outside the scope of validity of DIN VDE 0751. Siemens AG SPR2-250.831.01.03.02 ARCOSKOP Page 25 of 26 Medical Solutions 02.05...
Page 26
Changes to Previous Version Changes to Previous Version Converted documentation to DMS ARCOSKOP SPR2-250.831.01.03.02 Siemens AG Page 26 of 26 02.05 CS PS 24 Medical Solutions...