Once the cytochemical reactions are complete in the reaction chambers, the
sample and reagent mixtures from the perox, RBC, baso, and retic reaction
chambers are sent to the flowcells for analysis.
The Hgb reaction chamber serves as an optical cuvette through which the
hemoglobin measurement is read.
After analysis, the sample and reagent mixture is evacuated into the waste
container and the appropriate pathways and reaction chambers are rinsed.
Test results are sent to the computer to be reviewed and edited.
How the ADVIA 2120/2120i Software Works
The ADVIA 2120/2120i system software is a special-purpose program that runs
on the Windows NT operating system. You can navigate through the software
and operate the system using the mouse.
Status lines
There are two status lines that display messages about the system. Both lines
consist of two parts, each displaying different information.
First Status Line
Left Side:
Right Side:
1-16
The system saves all messages in the message log. Click the
message icon to view the last 10 messages.
Next sample ID, sample type, and test requested
Service, Notepad, Help, Printer, and Autoslide icons
Welcome to the ADVIA 2120/2120i Hematology System