Siemens ADVIA 2120 Operator's Manual page 243

Table of Contents

Advertisement

PEROX SHEATH Reagent Empty - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The reagent is empty, and the autosampler has
halted. The specified alarm criterion is user definable in the Customize Menu,
System Setup, Alarm/Stop Criteria, Reagent Conditions.
Corrective Action
The stop counter is automatically reset after reagents are replenished and the
Reagent Installation window is updated. Go to Logs, Reagent Log, Reagent
Installation. Select Start/Stop to continue the run.
PEROX SHEATH Reagent Expired
Present date is beyond the PEROX 3 SHEATH reagent expiration date, as
determined by encoded reagent container barcode label or as calculated from
installation date and open container stability.
Corrective Action
Replenish expired reagents and update the Reagent Installation window. Go to
Logs, Reagent Log, Reagent Installation.
PEROX SHEATH Reagent Low - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message, indicating that the reagent level is low. The
specified alarm criterion is user definable in the Customize Menu, System Setup,
Alarm/Stop Criteria, Reagent Conditions.
Corrective Action
The alarm counter is automatically reset after reagents are replenished and the
Reagent Installation window is updated. Go to Logs, Reagent Log, Reagent
Installation.
PEROX Temperature Out - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
7-74
Status Line messages