Autosampler - Autosampler is Already Running
The autosampler received a command to start or eject racks while it was running.
Autosampler - Bad Mixer Aspiration Cal
Mixer-aspiration calibration failed in diagnostic mode (Exerciser tab).
Call Siemens Service for assistance. Only Siemens Service personnel are
authorized to take these corrective actions.
Possible Cause
1
Autosampler
calibration
error.
2
Autosampler
CPU board
failed to store
calibration data
in flash
memory.
Autosampler - Bad Mixer Shuttle Cal
Mixer shuttle/index calibration failed in diagnostic mode (Exerciser tab).
Call Siemens Service for assistance. Only Siemens Service personnel are
authorized to take these corrective actions.
Possible Cause
1
Autosampler
calibration
error.
2
Autosampler
CPU board
failed to store
calibration data
in flash
memory.
Status Line messages
CAUTION
Corrective Action
Calibrate mixer aspirate position again.
Replace autosampler CPU board.
CAUTION
Corrective Action
Calibrate mixer shuttle/index position again.
Replace autosampler CPU board.