The default selectivity has been applied for the affected sample. Rerun the
sample if desired.
Cycle Aborted for Unknown Reason - Reset Required
A cycle abort condition was triggered by the analyzer, but the reason given was
not recognized by the computer.
Corrective Action
1
Reset the analyzer by selecting
Reset
2
After about 1 minute, if the status line does not display Ready to Run and
the green ready aspirate light is still off, follow these steps to exit the
ADVIA 2120/2120i software, turn off the analyzer, and then restart.
a
b
c
d
D
Data Manager Critical Error
An internal software communication error has triggered a high-level Data
Manager error.
Corrective Action
1
Select
2
Select
software.
If the problem persists, call Siemens Service for assistance.
Defined More Than Once In File
Note:
Possible Cause
1
The same keyword is set
several times in the header of
the Report.par, Prevalid.par,
Print.par, or QCreport.par
files. No blank or presentation
line can be inserted after the
asterisk (*) in the result block.
7-46
.
Select
Shut Down ADVIA
Select
at the analyzer touchpad.
Off
Select
CTRL+ALT+DELETE
software.
Restart the analyzer by selecting
Shut Down ADVIA
CTRL+ALT+DELETE
This message can begin with a keyword or a test code.
Utilities, Exerciser, Indicators, Analyzer
at the Log On/Off window.
and log back onto the system to restart the
at the analyzer touchpad.
On
at the Log On/Off window.
and log back onto the system to restart the
Corrective Action
Use the Format window to correct the file
layout.