Siemens ADVIA 2120 Operator's Manual page 188

Table of Contents

Advertisement

Error. Partial
Reset is
required.
2
Critical
Sampler
Error. Full
Reset is
required.
3
Air lines
supplying the
centering
collar are
pinched.
4
Faulty output-
queue sensor
or sensor
wiring.
Autosampler - Collar Failed to Extend, Reset Autosampler
The autosampler centering collar failed to extend properly.
Possible Cause
1
Critical
Sampler
Error. Partial
Reset is
required.
2
Critical
Sampler
Error. Full
Reset is
required.
Status Line messages
b
Close the autosampler access door. The autosampler
will reset.
a
At the analyzer touchpad, select
b
Wait about 1 minute and select
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
a
Open the autosampler access door.
b
Check for kinked or pinched tubing. Replace if
necessary.
c
Close the autosampler access door. The autosampler
will reset.
d
If the problem persists, call Siemens Service for
assistance.
Call Siemens Service for assistance.
Corrective Action
a
Open the autosampler access door.
b
Close the autosampler access door. The autosampler
will reset.
a
At the analyzer touchpad, select
b
Wait about 1 minute and select
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
.
Off
On
to restart the
Off
.
On
to restart the
7-19