Sample xxx Is Not Transmitted to the Host
There is a transmission problem to the host computer. The message specifies the
Sid# of the nontransmitted samples. Another message will follow identifying the
specific problem.
Scan Denied - Reset Autosampler
Barcode reader is unable to scan.
Possible Cause
1
Faulty barcode
reader cabling
2
Faulty barcode
reader
SD Not Defined For Test# xx Control# xxyyzzz
No standard deviation is defined for this test and for this control in the Control
Dictionary.
Corrective Action
Define the SD in the Control Dictionary.
Selected Printer Not Available - Result Report for SID#:
The printer is not responding to a print request.
Corrective Action
Check that the printer is powered on and has paper. Refer to the Message Log for
details.
Selector Valve Error
Potential software communication error or a hardware failure associated with the
selector-valve assembly.
Possible Cause
1
Hardware
failure
Status Line messages
Corrective Action
a
At the analyzer touchpad, select
b
Wait about 1 minute and select
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
See 1a, b, c
Corrective Action
a
Reset the analyzer by selecting
Indicators, Analyzer Reset
b
After about 1 minute, if the status line does not display
Off
.
to restart the
On
Utilities, Exerciser,
.
7-89