Siemens ADVIA 2120 Operator's Manual page 41

Table of Contents

Advertisement

To run samples from the autosampler:
1. Load samples in the following order:
a. Insert tube into rack with the barcode label visible above the rack
b. Load rack onto input queue with labels facing front of analyzer.
2. If the Standby indicator is lit, press Standby.
3. At the touchpad, select
The Start and Rack in Sampler indicators are lit.
4. Evaluate control results or validate patient results when available.
To run samples from the manual closed-tube sampler:
1. If the Standby indicator is lit, select
2. Run samples in the following order:
3. Scan the tube label or enter the sample information in the Manual Sample ID
tab.
4. Aspirate the sample.
a. Insert and push down tube containing the well-mixed sample into the
b. Sample is automatically aspirated – the sampling light flashes.
c. When the sampling light stops flashing, remove the tube.
5. Evaluate control results or validate patient results when available.
Daily Routine
Whole blood primer (primer label)
Controls (control label)
Patient samples (sample ID label)
barcode label that indicates the rack number and sample position. Do not
twist tube within rack.
Start/Stop Sampler
Whole blood primer (primer label)
Controls (control label)
Patient samples (sample ID label)
IMPORTANT:
Make sure the correct sample ID appears on the status line before
aspirating a sample using either the manual open-tube sampler or the manual
closed-tube sampler.
Waiting
for a matching workorder.
Make sure the correct sample selectivity appears on the status line. If you run a
control with a mismatched selectivity (Example: running a CBC/Diff control with
a Retic selectivity), the results will not appear in the Review /Edit tab and, in
some cases, the system computer may require a restart.
manual closed-tube sampler. Hold tube parallel to the sampler well wall.
.
.
Standby
displays on the status line while the system searches
3-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Advia 2120i

Table of Contents