Siemens ARCADIS Varic Replacement Instructions Manual

Siemens ARCADIS Varic Replacement Instructions Manual

Hide thumbs Also See for ARCADIS Varic:

Advertisement

Quick Links

ARCADIS
System Manual
Replacement Instructions
Replacement of the ON/OFF Assembly
These instructions are valid for the replacement of the ON/OFF
assembly, material no. 80 81 932, on ARCADIS Varic systems,
material no. 80 80 017 up to serial no. 10316, and
ARCADIS Orbic systems, material no. 80 81 080 up to
serial no. 20085.
n.a.
Print No.: SP00-000.841.01.01.02
Replaces: n.a.
SP
© Siemens AG 2005
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date:
08.05

Advertisement

Table of Contents
loading

Summary of Contents for Siemens ARCADIS Varic

  • Page 1 Replacement of the ON/OFF Assembly These instructions are valid for the replacement of the ON/OFF © Siemens AG 2005 assembly, material no. 80 81 932, on ARCADIS Varic systems, The reproduction, transmission or use of this document or its contents is not material no.
  • Page 2 0 - 2 Revision level Chapter Page Rev. 0 - 4 ARCADIS n.a. SP00-000.841.01 Page 2 of 4 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 3: Table Of Contents

    System leakage current measurement......3 - 9 ON/OFF assembly circuit diagram extract..... . . 3 - 10 Siemens AG n.a.
  • Page 4 0 - 4 Contents 4 ______ Changes to previous version _____________________________________4 - 1 ARCADIS n.a. SP00-000.841.01 Page 4 of 4 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 5: Notes And Symbols

    These instructions are valid for the replacement of the ON/OFF assembly, material no. 80 81 932 on ARCADIS Varic systems, material no. 80 80 017 up to serial no. 10316, ARCADIS Orbic systems, material no. 80 81 080 up to serial no. 20085.
  • Page 6: Safety Information

    CAUTION Electrical voltage! Non-compliance can lead to material damage. When working on the system, ESD regulations must be observed. ARCADIS n.a. SP00-000.841.01 Page 2 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 7: Radiation Safety Information

    Perform the required work with special care and attention to detail. If needed, wear working gloves. Siemens AG n.a. SP00-000.841.01 Page 3 of 10 ARCADIS Medical Solutions Rev.
  • Page 8: Safety Information - Risk Of Infection

    C-arm chassis and monitor trolley) must be moved section by section to detect broken conductors. The ground wire resistance must not exceed 0.2 ohms. ARCADIS n.a. SP00-000.841.01 Page 4 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 9 Fig. 1 Measuring circuit for measuring the ground wire resistance in systems that are disconnected from the power supply, according to DIN VDE 0751-1:2001-10, Fig. C2. 1 = System 2 = Applied part (not present) 3 = Measuring arrangement (integrated in the measuring instrument) Siemens AG n.a. SP00-000.841.01 Page 5 of 10 ARCADIS Medical Solutions Rev.
  • Page 10: Information On Measuring The System Leakage Current

    However, please note the minimum resolution of the system leakage current measuring instrument and any additional manufacturer's instructions restricting the use of the mea- suring device. ARCADIS n.a. SP00-000.841.01 Page 6 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 11 Fig. 3 Measuring circuit for the direct measurement of the system leakage current according to DIN VDE 0751-1:2001-10, Fig. C5 for protection class 1 = System 2 = Applied part (not present) 3 = Measuring arrangement (integrated in measuring instrument) Siemens AG n.a. SP00-000.841.01 Page 7 of 10 ARCADIS Medical Solutions Rev.
  • Page 12 The evaluation is not necessary in the case of a new determination. File the protocol sheet in the system folder or log book. ARCADIS n.a. SP00-000.841.01 Page 8 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 13: Required Documents

    General 1 - 9 Required documents ARCADIS Varic: General Safety Information TD00-000.860.01..Circuit Diagram G2266 Replacement Instructions SPR2-310.841.01..ARCADIS Orbic: General Safety Information TD00-000.860.01..Circuit Diagram G2266 Replacement Instructions SPR2-320.841.01..Required tools and measurement devices Standard tool kit* Digital multimeter e.g.
  • Page 14 1 - 10 General This page is intentionally left blank. ARCADIS n.a. SP00-000.841.01 Page 10 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 15: Opening The Monitor Trolley

    Remove the lower cover on the rear of the monitor trolley (Fig. 1). - The ON/OFF assembly and the uninterruptible power supply (UPS) are accessible. • Pull out the two power plugs on the UPS. Siemens AG n.a. SP00-000.841.01 Page 1 of 4...
  • Page 16 Loosen the plug connection X22 (2/Fig. 3). - The relay assembly is loose and can be removed. It is not replaced. ARCADIS n.a. SP00-000.841.01 Page 2 of 4 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 17 Pull the power cable and the monitor trolley cable out of the ON/OFF assembly. • The ON/OFF assembly can now be taken out of the monitor trolley. Due to the weight of the assembly, 2 persons are needed for lifting it. Siemens AG n.a. SP00-000.841.01 Page 3 of 4...
  • Page 18 2 - 4 Removing the ON/OFF Assembly This page is intentionally left blank. ARCADIS n.a. SP00-000.841.01 Page 4 of 4 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 19: Preparations

    For mains voltages of 200 V AC to 240 V AC, 15 A slow-blow fuses must be used. For mains voltages of 100 V AC to 127 V AC, 20 A slow-blow fuses must be used. Siemens AG n.a. SP00-000.841.01...
  • Page 20: Wiring

    3 - 2 Installing the ON/OFF assembly Wiring Layout D50 Fig. 3 ARCADIS n.a. SP00-000.841.01 Page 2 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 21: Connecting The Power Cable

    PC (imaging system) again, plug them into D66.X1 and D66.X5 and fasten them. Wiring of CAN cable (3D Reconstruction option) • If the 3D Reconstruction option is present, run the CAN connector to the CAN-to-USB converter again, plug it in and fasten it. Siemens AG n.a. SP00-000.841.01 Page 3 of 10 ARCADIS Medical Solutions Rev.
  • Page 22: Replacement And Wiring Of Ps/2 Cable

    • Plug in connector D50.X8. • Plug in connector D50.X6. • Plug in connector D50.X7. • Plug in connector D50.X13. ARCADIS n.a. SP00-000.841.01 Page 4 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 23: Wiring Of Terminals D50.X2, D50.X11 And D50.X4

    Insert the ON/OFF assembly into the monitor trolley and fasten it with the 4 Allen screws. In doing so, ensure that there is no mechanical strain on the cables and that they are not pinched. Siemens AG n.a. SP00-000.841.01 Page 5 of 10...
  • Page 24: Checking The Ground Wire Connections Of The On/Off Assembly

    C-arm chassis image intensifier housing Tab. 1 Connecting the UPS Connect the power input plug and power output plug to the UPS. ARCADIS n.a. SP00-000.841.01 Page 6 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 25: Startup

    On button. - The imaging system continues to shut down. - After that and without the On button being pressed again, the system restarts and is ready for operation after booting. Siemens AG n.a. SP00-000.841.01 Page 7 of 10 ARCADIS Medical Solutions Rev.
  • Page 26: Concluding Work

    0.2 ohms is not exceeded, indicates errors in the ground wire connections. (Ground wire or contacts) The ground wire resistance must not exceed 0.2 ohms. ARCADIS n.a. SP00-000.841.01 Page 8 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 27: System Leakage Current Measurement

    2.5 mA is not exceeded. NOTE Where DIN VDE 0751-1:2001-10 applies, the permissible system leakage current limit value is 2.5 mA. In addition, the national regulations must be observed. Siemens AG n.a. SP00-000.841.01 Page 9 of 10 ARCADIS Medical Solutions Rev.
  • Page 28: On/Off Assembly Circuit Diagram Extract

    3 - 10 Installing the ON/OFF assembly ON/OFF assembly circuit diagram extract Fig. 4 ARCADIS n.a. SP00-000.841.01 Page 10 of 10 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...
  • Page 29 Changes to previous version 4 - 1 New publication Siemens AG n.a. SP00-000.841.01 Page 1 of 2 ARCADIS Medical Solutions Rev. 01 08.05 CS PS 24 System Manual...
  • Page 30 4 - 2 Changes to previous version This page is intentionally left blank. ARCADIS n.a. SP00-000.841.01 Page 2 of 2 Siemens AG System Manual Rev. 01 08.05 CS PS 24 Medical Solutions...