placed,
obscured, or
not present.
2
Barcode reader
faulty.
Sample Rejected: SID #
The results for the last aspiration of this sample have been rejected. This sample
record is full. A sample record can hold only 110 results, including C.Res.
Prev.Run, and Prev.Res.
Corrective Action
Create a new workorder with the same patient demographics and assign a new
SID #. Label the sample tube again with the new SID #, and run.
Sample Result Data Communication Failure
The Align Optics or Direct Optics cycle was exited before result processing was
complete.
Corrective Action
Results produced when this error was triggered are invalid. Repeat sample
aspiration for the affected samples.
Sample SID Not Found - Printing Aborted
The sample displayed in the Review and Edit window does not exist in the
database. The system cannot provide the requested printout.
Corrective Action
Correct the problem using the Key list in the Format window.
Sample Status: – THE STATUS – Sid#:x R&P#:y
A sample with this status could not receive results. The results received for this
sample have been refused.
Corrective Action
In the Customer Parameters window, modify the parameter that specifies each
sample status allowed to receive results.
7-88
the stopper). Verify that the barcode is clean and of
specified quality.
Call Siemens Service for assistance.