Scrambler Board, or
Can Bus Scrambler
Board.
SHEATH / RINSE Reagent Container Empty
SHEATH/RINSE level has dropped below the level sensor in the
SHEATH/RINSE container, indicating that the container is empty. When this
error message is displayed, the aspirate plate will not function, nor will the Ready
light illuminate.
Possible Cause
1
SHEATH/RINSE
is empty.
2
Faulty level
sensor.
SHEATH / RINSE Reagent Expired
Present date is beyond the SHEATH/RINSE reagent expiration date, as
determined by encoded reagent container barcode label or as calculated from
installation date and open container stability.
Corrective Action
Replenish expired reagents and update the Reagent Installation window. Go to
Logs, Reagent Log, Reagent Installation.
Sid# Already Exists - Result Message Rejected
You have created a workorder with an Sid# having the format: 999XXXX. This
format is used by the analyzer to send results of misread samples. The results are
refused.
Corrective Action
Create a workorder with a proper Sid# format.
Standby
The analyzer is in a low-power state that reduces component wear during periods
of inactivity. The green ready-to-aspirate light is off. Typically, the analyzer
enters standby automatically. Use the touchpad Standby button to exit or enter
the standby state manually.
Corrective Action
Status Line messages
Corrective Action
Replace SHEATH/RINSE reagent and use the message
icon to Remove the error message.
Call Siemens Service for assistance.