2. Open the optics cover. The perox flowcell is on the left.
Perox flowcell (
3. Disconnect the shuttle line (
sheath-stream input line (
4. Locate hydraulic valve # 24, Unscrew the threaded fitting from the top right
side of the valve.
5. Release the flowcell by loosening the release knob located on the right of the
flowcell adjustment assembly
5-40
), RBC flowcell (
)
1
2
1
), the sample-stream input line (
) from the CFM. Allow these lines to hang freely.
3
1 SHEATH NIPPLE
2 SAMPLE NIPPLE
3 SHUTTLE NIPPLE
2
), and the
Troubleshooting the Analyzer