Siemens ADVIA 2120 Operator's Manual page 173

Table of Contents

Advertisement

Possible Cause
1
PSI pressure
gauge is out of
range
2
Pressure leak
20 PSI Out of Range - Stop
System pressure is out of range. The system has stopped because the pressure has
reached a critical level that may affect the integrity of sample results and damage
the system.
Possible Cause
1
PSI pressure
gauge is out of
range
2
Pressure leak
IMPORTANT
If problems persist, Call Siemens Service for assistance. Further corrective action
must be taken by Siemens service personnel only.
4
4 Key Not Found
The Control Dictionary or the Alarm Dictionary is empty. If the Alarm
Dictionary is empty, sample/system flags will appear as exclamation points (!).
Possible Cause
1
The Data Manager
received a sample that is
identified as a control but
that has not been set up in
the Control Dictionary.
2
The Alarm Dictionary is
empty or missing the Delta
Check alarm. The Delta
Check alarm must be in
the Alarm Dictionary
even if you do not use
delta checking.
7-4
Corrective Action
Check the 20 PSI gauge and adjust the regulator if
necessary.
±
Setpoint: 20 PSI
1.
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
Corrective Action
Check the 20 PSI gauge and adjust the regulator if
necessary.
±
Setpoint: 20 PSI
1.
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
Corrective Action
Do not use control ID labels on samples, or
set up the control in the Control
Dictionary.
Enter alarms in the Alarm Dictionary.
Status Line messages